欢迎访问《茶叶科学》,今天是

茶叶科学 ›› 2008, Vol. 28 ›› Issue (2): 93-100.doi: 10.13305/j.cnki.jts.2008.02.005

• • 上一篇    下一篇

丝网印刷碳电极传感器检测茶叶中痕量铅研究

赵广英, 吴艳燕   

  1. 浙江工商大学食品与生物工程学院,浙江省食品安全重点实验室 杭州 310035
  • 收稿日期:2007-07-25 修回日期:2007-12-07 出版日期:2008-04-15 发布日期:2019-09-16
  • 作者简介:赵广英(1960–),女,教授,研究方向为食品质量安全快速检测,食品卫生微生物检验,E-mail:zhaogy-user@163.com
  • 基金资助:
    国家自然科学基金(No.30571623); 浙江省重中之重学科(No. Z05-29); 浙江省食品质量安全重点实验室开放项目(No. Z04-179)

Determination of Trace Lead in Tea with Screen-printed Carbon Electrodes

ZHAO Guang-ying, WU Yan-yan   

  1. Food safety key lab of Zhejiang, College of Food Science and Biotechnology Engineering, Zhejiang Gongshang University, Hangzhou 310035
  • Received:2007-07-25 Revised:2007-12-07 Online:2008-04-15 Published:2019-09-16

摘要: 建立了同位镀汞法修饰的丝网印刷碳电极电化学方波溶出伏安法快速检测茶叶中的铅的方法。详细描述了建立该系统检测方法的条件优化过程,得到了一系列最佳参数:最佳的镀汞液浓度4×10-4 mol/L,沉积电位-1.1 V,电位增量3 mV,方波频率10 Hz,方波幅度0.05 V,pH值5,沉积时间280 s,平衡时间30 s。在优选条件下获得的方法灵敏度、线性范围和检测限分别为22.7 nA/(μg/L),10~225 μg/L (r=0.9986)和0.74 μg/L(S/N=3)。研究了对多种重金属元素存在条件下120 μg/L的铅检测的干扰情况。本试验对样品前处理方法研究发现,采用家用微波炉、聚四氟乙烯密封增压微波消化罐消化茶叶,用标准加入法对分析样品进行检测,与国标方法相比较,无显著偏差。该方法灵敏、准确、快速适合于茶叶中痕量铅的测定。

关键词: 方波溶出伏安法, 丝网印刷碳电极, 茶叶

Abstract: The square wave anodic stripping voltammetric (SWASV) methodology using a screen printed carbon electrode (SPCE) in situ plated in mercury (Ⅱ) was developed by the authors to determine lead in tea. A detailed description of the systematic laboratory test carried out to optimize its performance is reported. A series of optimum parameter are as follows Hg2+concentration: 4×10-4 mol/L, Plating potential: -1.1 V, Estep: 3 mV, Frequency: 10 Hz, Amplitude: 60 mV, pH: 5, plating time: 280 s, balance time: 30 s. Under the optimized analytical condition, the obtained sensitivity, linearity, and detection limit are 22.7 nA/(μg/L), 10~225 μg/L (r=0.9986), and 0.74 μg/L(S/N=3),respectively. The interference effect was thoroughly studied with various metals in the determination of 120 μg/L Pb2+. Tea was digested using domestic microwave oven and PTFE sealed and pressurized microwave digestion tank. The standard method of analysis was used to detect the samples and comprised with the GB method, no significant deviation was found. The method was sensitive, accurate, fast and suitable for the determination of trace lead in tea.

Key words: square wave anodic stripping voltammetry, screen printed carbon electrode, tea

中图分类号: